Skip to main content
menu

Camilla - BD LSRFortessa - FACSDiva based

MIFLOWCYT information specific for this instrument can be found in the FCC_Library.

Detailed information about the QC can be found on the baseline report (available upon request).

Reservations can be made online by following this link.

You will need an approved USER account. Apply for one here.

To design a panel using Fluorofinder please click here.

Training videos can be accessed here.

LSR Fortessa Configuration: Low starting voltages as of Jan 2023 posted at the machine and available via FCC_Library.

BLUE 488 LASER (50mW):

Detector

LP

Band Pass

Effective Ranges (nm)

Fluor's (check Fluorofinder for more info)

BLUE A

685

710/50

685-735

PerCP-Cy5.5

BLUE B

505

530/30

515-545

BB 515

BLUE C

  488/10   SSC

FSC

 

 

 

 

RED 640 LASER (40mW):

Detector

LP

Band Pass

Effective Range (nm)

Fluor's (check Fluorofinder for more info)

RED A

750

780/60

750-810

APC-Cy7

RED B

690 730/45 707.5-750 APC-R700

RED C

 

670/14

663-677

APC

VIOLET 405 LASER (50mW):

Detector

LP

Band Pass

Effective Range (nm)

Fluor's (check Fluorofinder for more info)

VIOLET A

750

780/60

750-810

BV 786

VIOLET B

690

710/50

690-735

BV 711

VIOLET C

630

660/20

650-670

BV 650

VIOLET D 600 610/20 600-620 BV 605
VIOLET E 505 525/50 505-550 BV 480
VIOLET F   431/28 417-445 BV 421

YELLOW-GREEN 561 LASER (50mW):

Detector

LP

Band Pass

Effective Range (nm)

Fluor's (check Fluorofinder for more info)

YG A

750

780/60

750-810

PE-Cy7

YG B

685

710/50

685-735

PE-Cy5.5

YG C

635 670/30 655-685 PE-Cy5

YG D

600

610/20

600-620

PE-Texas Red

YG E

 

586/15

578.5-593.5

PE, PI, Alexa 532

Ultra Violet 355 LASER (20mW):

Detector

LP

Band Pass

Effective Range (nm)

Fluor's (check Fluorofinder for more info)

UV A

690

740/35

722.5-757.5

BUV 737

UV B

 

379/28

365-393

BUV 395