Camilla - BD LSRFortessa - FACSDiva based
MIFLOWCYT information specific for this instrument can be found in the FCC_Library.
Detailed information about the QC can be found on the baseline report (available upon request).
Reservations can be made online by following this link.
You will need an approved USER account. Apply for one here.
To design a panel using Fluorofinder please click here.
Training videos can be accessed here.
LSR Fortessa Configuration: Low starting voltages as of Jan 2023 posted at the machine and available via FCC_Library.
BLUE 488 LASER (50mW):
Detector
|
LP
|
Band Pass
|
Effective Ranges (nm) |
Fluor's (check Fluorofinder for more info)
|
BLUE A
|
685
|
710/50
|
685-735 |
PerCP-Cy5.5
|
BLUE B
|
505
|
530/30
|
515-545 |
BB 515
|
BLUE C
|
|
488/10 |
|
SSC |
FSC
|
|
|
|
|
|
RED 640 LASER (40mW):
Detector
|
LP
|
Band Pass
|
Effective Range (nm) |
Fluor's (check Fluorofinder for more info)
|
RED A
|
750
|
780/60
|
750-810 |
APC-Cy7
|
RED B
|
690 |
730/45 |
707.5-750 |
APC-R700 |
RED C
|
|
670/14
|
663-677 |
APC
|
|
VIOLET 405 LASER (50mW):
Detector
|
LP
|
Band Pass
|
Effective Range (nm) |
Fluor's (check Fluorofinder for more info)
|
VIOLET A
|
750
|
780/60
|
750-810 |
BV 786
|
VIOLET B
|
690
|
710/50
|
690-735 |
BV 711
|
VIOLET C
|
630
|
660/20
|
650-670 |
BV 650
|
VIOLET D |
600 |
610/20 |
600-620 |
BV 605 |
VIOLET E |
505 |
525/50 |
505-550 |
BV 480 |
VIOLET F |
|
431/28 |
417-445 |
BV 421 |
|
YELLOW-GREEN 561 LASER (50mW):
Detector
|
LP
|
Band Pass
|
Effective Range (nm) |
Fluor's (check Fluorofinder for more info)
|
YG A
|
750
|
780/60
|
750-810 |
PE-Cy7
|
YG B
|
685
|
710/50
|
685-735 |
PE-Cy5.5
|
YG C
|
635 |
670/30 |
655-685 |
PE-Cy5 |
YG D
|
600
|
610/20
|
600-620 |
PE-Texas Red
|
YG E
|
|
586/15
|
578.5-593.5 |
PE, PI, Alexa 532
|
Ultra Violet 355 LASER (20mW):
Detector
|
LP
|
Band Pass
|
Effective Range (nm) |
Fluor's (check Fluorofinder for more info)
|
UV A
|
690
|
740/35
|
722.5-757.5 |
BUV 737
|
UV B
|
|
379/28
|
365-393 |
BUV 395
|
|