Zoot - BD Symphony A1 - FACSDiva based
MIFLOWCYT information specific for this instrument can be found in the FCC_Library.
Detailed information about the QC can be found in the baseline report (available upon request).
Reservations can be made online by following this link.
You will need an approved USER account. Apply for one here.
To design a panel using Fluorofinder please click here.
Training Videos can be accessed here.
Configuration: Low starting voltages as of Jan 2023 posted at the machine and available in FCC_Library.
BLUE 488 LASER (100mW):
Detector
|
LP
|
Band Pass
|
Fluor's (check Fluorofinder for more info)
|
BLUE B
|
690
|
710/50
|
PerCP-Cy5.5
|
BLUE C
|
505
|
530/30
|
FITC, GFP, Alexa 488
|
BLUE E
|
|
488/10 |
SSC |
BLUE F |
|
488/10 |
Small Particle SSC |
FSC
|
|
|
|
|
RED 637 LASER (100mW):
Detector
|
LP
|
Band Pass
|
Fluor's (check Fluorofinder for more info)
|
RED A
|
750
|
780/60
|
APC-Cy7, APC-Alexa 750
|
RED B
|
690 |
710/50 |
APC-Cy5.5, Alexa 700, Alexa 680 |
RED C
|
655
|
670/30
|
APC, Alexa 647
|
|
VIOLET 405 LASER (100mW):
Detector
|
LP
|
Band Pass
|
Fluor's (check Fluorofinder for more info)
|
VIOLET A
|
750
|
780/60
|
BV 786
|
VIOLET B
|
690
|
710/50
|
BV 711
|
VIOLET C
|
655
|
670/30
|
BV 650
|
VIOLET D |
595 |
610/20 |
BV 605 |
VIOLET E |
505 |
525/50 |
Pacific Orange, BV 510, |
VIOLET F |
410 |
431/28 |
Pacific Blue, BV 421 |
|
YELLOW-GREEN 561 LASER (100mW):
Detector
|
LP
|
Band Pass
|
Fluor's (check Fluorofinder for more info)
|
YG A
|
750
|
780/60
|
PE-Cy7, PE-Alexa 750
|
YG B
|
690
|
710/50
|
PE-Cy5.5
|
YG C
|
655 |
670/30 |
7AAD, PE-Cy5 |
YG D
|
595
|
610/20
|
PE-Alexa 610, PE-Texas Red
|
YG E
|
580
|
586/15
|
PE, Alexa 532
|
|