Skip to main content
menu

Mass Cytometry

The Helios Mass Cytometer (known as Ludo) is the latest in technology and high end instrumentation from Fluidigm (formally DVS Sciences). This instrument functions similar to traditional flow cytometry except for using metal in place of fluorescence to tag antibodies. With the use of metal-ion tagged antibodies, extremely high multiparametric analysis (30-40+ simultaneous parameters) can be done on your samples.

Mass Cytometry

For a general overview of mass cytometry, please read this presentation. There is additional info in the Library share of the server for anyone interested in more information. Go to the Registration & Training page for instructions on how to connect.

General Information before running samples

  1. Before samples can be submitted to the Flow Core for analysis, an initial meeting including the PI and members of the PI’s lab who will be doing the labeling and the FCR is required. This will ensure we have the best success with adopting this new technology. Times for meeting can be made by appointment. Please contact Matt Cochran if you'd like to schedule a time.
  2. Review the reagents listed/available at the Fluidigm website.
  3. Labeling Protocols - Protocols for labeling are in the Flow Core Server Library.
  4. Data Analysis in mass cytometry experiments can be performed using any flow cytometry analysis software using traditional gating methods. In addition, most analysis packages including FCS Express and Flowjo now include additional tools to deal with the highly multiparametric data from mass cytometry panels.  For a discussion of how the Flow Core can assist in your data analysis needs please contact  Matt Cochran.
  5. Sample Preparation - There is specific information a user needs to be aware of when planning how samples need to be submitted for analysis
    1. Sample Concentration - samples need to be no more than one million cells per ml. If the concentration is higher, data quality will suffer.
    2. The sample can be run via the Supersampler (for larger continuous volume samples that need to be kept cold) or via the PSI unit for continuous stable samples.
    3. Sample Running Duration - The instrument has a fixed sample introduction rate of 45uL/min meaning that 1mL of sample will run for ~23 minutes via the Supersampler.
      1. Wash cycles between samples vary (3-10 minutes).
    4. Barcoding of samples - Samples can be batched and barcoded via reagents available through Fluidigm which will reduce staining variation, improve throughput on the instrument, and reduce any contribution of instrument variations.  For these reasons we recommend barcoding of samples when possible.

Helios Submission Form

A Helios Acquisition Form is required prior to a reservation being confirmed on the calendar.  A copy of  the form can be accessed here.

Please allow a minimum of 72 hours advanced notice. The staff will do their best to accommodate all requests but do not begin the experiment until a time is officially scheduled.

The following information should be taken into account when considering/planning a mass cytometry experiment:

  • Concentration of samples should not be more than 1 million/mL. Data quality can suffer at higher concentrations.
  • Know your sample volumes and how you would like your sample to be introduced to the instrument.
  • How many events do you need for analysis? Average recovery is ~50% of sample input.
  • Know your panel: Be prepared to double check the template.
  • Most samples must be washed and resuspended in ultrapure water.
  • All samples will be filtered by FCC staff prior to the run.
  • Late arrival of samples could result in having to postpone the run and impact data quality.
  • Bring samples in locking secondary containment with biohazard label.

If you have trouble filling out the form, please send us an email or call 276-5330 and ask for Zach.

Thank you,
The CyTOF Team

Terry 
Zach 
Steve