Configuration:

UV 349 Laser (60mW)

Detector LP BP Fluor's
UV A 770 820/60 BUV 805
UV B 690 740/35 BUV 737
UV C 635 670/30 BUV 661
UV D 550 586/15 BUV 563
UV E 450 515/30 BUV 496
UV F   379/28 BUV 395

BLUE 488 LASER (100mW):

Detector

LP

Band Pass

Fluor's

BLUE A

770

780/60

RB 780

BLUE B

690

710/50

RB 705

BLUE C

600 610/20 RB 613
Blue D 505 515/20 FITC
Blue E   488/10 SSC

FSC

 

 

 

RED 637 LASER (140mW):

Detector

LP

Band Pass

Fluor's

RED A

750

780/60

APC-Cy7

RED B

690 710/50 APC-R700

RED C

650

670/30

APC, Alexa 647

VIOLET 405 LASER (100mW):

Detector

LP

Band Pass

Fluor's

VIOLET A

750

780/60

BV786

VIOLET B

690

710/50

BV711

VIOLET C

635

670/30

BV650

VIOLET D 600 610/20 BV605
VIOLET E 450 470/14 BV480
VIOLET F 410 431/28 BV421

YELLOW GREEN 561 LASER (150mW):

Detector

LP

Band Pass

Fluor's

YG A

750

780/60

PE-Cy7

YG B

690

710/50

PE-Cy5.5

YG C

650 670/30 PE-Cy5

YG D

600

610/20

PE-Alexa 610

YG E

570

586/15

PE